Quality and verification of analytical methods of the renal and lipid profile in a Private Laboratory - Jipijapa 2024

Authors

DOI:

https://doi.org/10.62305/biosana.v5i1.441

Keywords:

Grubbs limits; precisión; coefficients of variation; veracity; bias

Abstract

Quality control is one of the most important impacts in laboratory testing, ensuring both the precision and accuracy of patient sample results. This study aimed to evaluate the quality and verification of analytical methods of the renal and lipid profile in the private laboratory-Jipijapa. A cross-sectional, prospective observational research study focused on analyzing data from different variables was applied. To carry out this research, we worked with normal and pathological controls from the TruLab brand and renal and lipid profile reagents from the DiaSys brand. Where 25 measurements were obtained, including 5 replicates in 5 days for each level evaluated. The results obtained for the verification and precision of each of the tests were accepted according to the statistical analysis carried out, in which the assigned concentration is within the limits where each of the results of the normal and pathological control are within the range. from the manufacturer. It is concluded that it is important to apply this type of method in laboratories, in order to obtain precise and reliable results, ensuring the quality and reliability of clinical analyses.

Downloads

Download data is not yet available.

References

Organización Panamericana de la Salud/Organización Mundial de la Salud. OPS. [Online]. [cited 2024 06 19. Available from: https://www.paho.org/es/temas/sistemas-laboratorio .

Cruz S, Bozo M, Molero T, Gómez M. Desempeño analítico en la determinación de colesterol y triglicéridos en laboratorios clínicos de la ciudad de Maracaibo, Venezuela. Revista Biomedicina. 204; 26(2).

Comité Técnico ISO/TC 212. Ensayos clínicos de laboratorio y sistemas de análisis de diagnóstico. Pubmed. 2009.

Gerarda Rodríguez-Benavides-Rigoberto Blanco-Sáenz. Aseguramiento de la calidad analítica y norma ISO 17 025. SCielo. 2001 Junio; 22(1-2).

Carmen Ricós,Pilar Fernandez-Calle, Maria Carmen Perich. Control interno de la calidad – bases del pasado, situación presente y futuras tendencias. Pubmed. 2022 Octubre; 3(3).

Pinares Astete FE, Cieza Zevallos JA, Castillo Fernández WN. Confiabilidad de las mediciones de urea e implicancias en el cálculo del Kt/V en pacientes en hemodiálisis. Scielo. 2020 Enero; 14(1).

Alberto Vílchez Rodríguez JGC,SEP,CVO,PLEB. Estudio de verificación de precisión y veracidad en un sistema Atellica. Pub Med. 2020 Septiembre; 1(4).

Solbellys Cruz MBTMMGMZAP. Desempeño analítico en la determinación de colesterol y triglicéridos en laboratorios clínicos de la ciudad de Maracaibo, Venezuela. Scielo. 2014 Junio ; 26(2).

Published

2025-02-01

How to Cite

Eulalia Adalay , D. A. ., Pillasagua Mero , A. R., & Lino Villacreses, W. A. (2025). Quality and verification of analytical methods of the renal and lipid profile in a Private Laboratory - Jipijapa 2024 . BIOSANA Health Scientific Journal. ISSN 2960-8481, 5(1), 281–300. https://doi.org/10.62305/biosana.v5i1.441